Bendtsen Type Tester Manual
-
Noviprofibre
-
Made in France
To measure the air permeability and surface roughness according to Bendtsen
Product description
- Surface roughness :
A sample is placed on the planar glass plate after starting the device. If needed, it can be flattened by a planar weight. The roughness head is placed carefully onto the sample.
The pressure is controlled by the rotating manostat weight. Values are indicated at the manometer tubes. Attention should be paid to the displayed value : only values within 20-
80 % are valid. - Air permeability :
The operator inserts the porosity head after turning on the compressor. The porosity head is clamped manually until the end of the test. By manual switching, the range between the
individual rotameter tubes can be chosen.
Technical specifications
General data
- The standard measurement pressure is 1.47 ± 0.02 kPa in accordance with the Standards.
- Two other pressures can be selected by changing the weight of the manostats (0.74 ± 0.01 kPa and 2.20 ± 0.03 kPa).
- Choice of three rotameter tubes to analyze a wide range of products.
- 5 - 150 ml per minute
- 50 - 500 ml per minute
- 300 - 3000 ml per minute
Log in
Electricity : 230 V, 50 Hz
Weight and dimensions Packaging
On request
Product weight and dimensions
10 Kg
500 x 500 x 500 mm ( WxLxH )
Scope of delivery
- Bendtsen Type Tester Manual
- Compressor
- Roughness head
- Porosity head
- Glass plate
- Operating manual
Interested in this product ?
Do not hesitate to ask for a quote
Related products
-
SpectrophotometerN5900 • N5950BISO 2469 • ISO 2470 • ISO 3688 • ISO 11475 • ISO 11476 • TAPPI T 519 • TAPPI T 525 • TAPPI T 527
Device for the determination of ISO brightness, color, color differences, fluorescence and opacity of paper.
-
Computerized friction coefficient equipment140104TAPPI T503 • T816 • T549
Device for determining the coefficient of static and dynamic computerized friction.
-
2D Formation Sensor device
Device for measuring the formation of the paper sheet.